Specifications


 
 
Instrument Configurations
:
P/N 6200 : 2-Plex (Green, Yellow)
P/N 62H0 : 2-Plex (Green, Yellow)+HRM (High Resolution Melt)
P/N 6500 : 5-Plex (Green, Yellow, Orange, Red, Crimson)
P/N 65H0 : 5-Plex (Green, Yellow, Orange, Red, Crimson)+HRM (High Resoultion Melt)
P/N 6600 : 6-Plex (Blue, Green, Yellow, Orange, Red, Crimson)

Rotor Configurations

:
30 µL x 100-wells: GeneDisc™ 100 (heat-sealed plate) AVAILABLE SOON
0.1 mL x 72-wells: GeneDisc™ 72 (heat-sealed plate)
0.1 mL x 72-wells: 0.1 mL strip tubes (strips of 4, matching caps)
0.2 mL x 36-wells: 0.2 mL tube (attached cap, domed or flat) 

Channel Information

:
Channel
Excite
Detect
Some Example Fluorophores Detected
Blue
365±20      
460±15 
Biosearch Blue™, Marina Blue®, Bothell Blue®, Alexa Fluor® 350
Green
470±10
510±5 
FAM™, SYBR® Green 1, Fluorescein, EvaGreen™, Alexa Fluor® 488
Yellow
530±5
555±5
JOE™, VIC™, HEX™, TET™, CAL Fluor® Gold 540, Yakima Yellow®
Orange
585±5
610±5
ROX™, CAL Fluor® Red 610, Cy3.5™, Texas Red®, Alexa Fluor®568
Red
625±10
660±10
Cy5™, Quasar™ 670, LightCycler Red640®, Alexa Fluor 633
Crimson
680±5
>712(LP)
Quasar™ 705, LightCycler Red705®, Alexa Fluor® 680
HRM
460±15
510±5
SYTO®9, LC Green®, LC Green® Plus+, EvaGreen™...

Thermal Performance

:

Uniformity: ± 0.01°C
Resolution: ± 0.02°C
Range: ambient–99°C
Maximum Ramp Rate: 10°C/sec (air)

Optical System

:

Fixed path length
Separate color high-intensity LED (light-emitting diode) excitation source per channel
Separate emission filter per channel

Detector

:

Photomultiplier (PMT) detector with variable or automatic gain setting (sensitivity control), user selected

Dimensions

:

370mm (14.6") wide, 420mm (16.5") deep, 560mm (22") deep door open, 275mm (10.8") high

Weight

:

14 kg (31 lbs)

Electrical
:

100-120VAC @ 60Hz, 200-240VAC @ 50Hz
Power consumption: 8VA (idle/standby), 560VA (peak) upgrades

Minimum Computer
:

Pentium™ IV, 2.6GHz (Desktop), Pentium M, 1.6GHz (Laptop) or equivalents
256Mb RAM, Windows XP®, USB or RS232 Serial Port connection 

Software
:
Extensive analysis, graphing and statistical functions built-in. Unlimited use software licence included, free upgrades (by web download), Windows XP, Pentium IV (2GHz) or higher PC, USB or Serial Port required

40 Cycle Run Time

:

Typically 40 minutes to 1.5 hours (rotor and protocol dependant)

Supported Volumes

:

5 µL to 100 µL (rotor and protocol dependant)
Typical reaction volume 20 µL

Instrument Color

:

Outback Red or Bondi Blue

HRM (High Resolution Melt)

:
New HRM capability is optional and comprises a dedicated high-performance optical subsystem, additional hardware and specific HRM analysis software (fully integrated). Thermal uniformity ±0.01°C, Resolution ±0.02°C, Range ambient–99°C. HRM data acquisition (read) rate of 20 reads for each 0.02°C increment (=1000 reads/°C)

NOTE: Conventional DNA Melt/Anneal capability and analysis software (including auto-call genotyping capability) is standard on all models
DNA Concentration Measurement
:

For direct measurement of nucleic acid concentration using fluorescent dyes (e.g. PicoGreen®, RiboGreen®, etc). Comprises dedicated analysis software (standard on all models)

 
 
 