The NanoDrop ® ND-1000 Spectrophotometer can be used to simultaneously measure the concentrations of proteins (A280) and up to two fluorescent dyes. In addition to measuring both components of a protein-dye conjugate, the versatile Proteins and Labels module can also be used to measure the purity of metalloproteins (such as hemoglobin) using wavelength ratios.

 
 


Key Features

Six different reference types available : 1 abs=1mg/ml, BSA, IgG, Lysozyme, user entered mass extinction coefficient, user entered molar extinction coefficient .
Measurement of wide variety of fluorescent dyes possible: Cy3, Cy5, Cy3.5, Cy5.5, Alexas 488, 546, 555, 594, 647 and 660
Concentration of nucleic acid and up to two dyes calculated and displayed
Reblank feature allows for the re-analysis of previous sample with a new blank
Full spectrum data is automatically archived in a text delimited file on the controller PC and easily reviewed using the Data Viewer feature
Sample ID field is key-focused for easy bar coder use
Absorbance is represented as a 1mm path equivalent

Measurement Concentration Range


Sample Type
Detection Limit
(pmol/ul)
Approximate Upper Limit (pmol/ul)
Typical Reproducibility
(minimum 5 replicates)
(SD= pmol/ul; CV= %)

Cy3, Cy3.5, Alexa Fluor 555, Alexa Fluor 660

0.20
100
sample range 0.20-4.0 pmol/ul: ± 0.20 pmol/ul
sample range >4.0 pmol/ul: ± 2%

Cy5, Cy5.5 and Alexa Fluor 647

0.12
60
sample range 0.12-2.4 pmol/ul: ± 0.12 pmol/ul
sample range >2.4 pmol/ul: ± 2%

Alexa Fluor 488 and Alexa Fluor 594

0.40
215
sample range 0.40-8.0 pmol/ul: ± 0.40 pmol/ul
sample range >8.0 pmol/ul: ± 2%

Alexa Fluor 546

0.30
145
sample range 0.30-6.0 pmol/ul: ± 0.30 pmol/ul
sample range >6.0 pmol/ul: ± 2%



 