Consort
 

Electrophoresis Power Supply

 

Manual programming:
 
Allows setting voltage, current, power and timing for a routine electrophoresis run. Parameters can be changed temporarily without interrupting the run.
 
Timer :
 Timer operation is a useful standard feature. The micro-computer will automatically terminate the run and sound an alarm when the count down of the selected value is achieved.
 
Automatic cross-over:
 Has constant voltage, constant current, constant power capabilities with automatic cross-over and shows which parameter is kept constant.
 
Automatic recovery after power failure:
 After a mains power failure the instrument will automatically continue the run for the remaining time.
 
Safety precautions: 
  • The user is protected from potential shock hazard since the AC lines is automatically disconnected from the high voltage transformer when a ground leakage path is detected.

  • The instrument is fully protected against any overload condition accidental short circuit of the output.

  • The high voltage cannot suddenly appear at the outputs. It will always increase smoothly upto one of the pre-set limits is reached.

 
SpecificationsOrdering InformationDownload

VOLTAGE
0...200 V
CURRENT
0...200 mA
POWER
0...20 W
PARAMETER RANGE
1...100% of full scale
TIMER
0...99:59 h
DISPLAY LCD
2x16 characters
RESOLUTION
1 V, 1 mA, 1 W
PROGRAMS
1 set of parameters
OUTPUTS
3 in parallel, 4 mm sock
MIN. LOAD RESISTANCE
15 Ω
NO LOAD DETECTION
on/off, programmable
GROUND LEAKAGE DETECTION
Yes
OVERLOAD DETECTION
Yes
AMBIENT TEMPERATURE
0...40°C
RELATIVE HUMIDITY
0...95%, non condensing
POWER REQUIREMENTS
210...250 VAC, 50/60 Hz
DIMENSIONS (WxDxH)
24x20x13 cm
WEIGHT
2 kg